4-14 Chapter 4
!
If the calibration was performed
for a multi-scale option, you must
enter a serial number. Key in the
serial number found on the option
board and press [ENTER]. This
number is stored in P61122.
GSE Scale Systems
4.
Restore power and enter the calibration routine:
100 [SELECT] 54321 [ID] [ENTER]
New Zero?
5.
At the
6.
The display prompts
7.
The display prompts
made the connections in step 3, press [ENTER].
8.
The display will show
calibration values for several seconds.
9.
The display prompts
10.
Remove the connections from step 3 which short the signal, sense
and shield.
11.
Attach a precision load cell simulator to the J1 load cell connector
using excitation and signal connections.
12.
Move the sense jumpers E1 & E2 back to the internal (INT) position.
13.
Set the simulator to 0 mV and press [ENTER].
14.
The display will show
calibration values for several seconds.
15.
The display prompts
and press [ENTER].
16.
The display shows
calibration values for several seconds.
17.
The display shows
complete. Press [ENTER].
18.
Press [ENTER] at the
ENTER=EXIT
prompt to save the new calibration and exit the
calibration mode.
- or -
To exit the calibration mode without saving the new calibration,
press [CLR] at the
ENTER=UNDO
the
to exit the calibration mode.
prompt, key in 23640 [ENTER].
ReCAL A/D?
. Press [ENTER].
Gnd. Input
. Assuming you have already
Mot'n Delay
and begin processing
Undo GND
Set To 0 mV
, then
Mot'n Delay
and begin processing
Set To 1 mV
. Set the simulator to 1 mV
Mot'n Delay
and begin processing
A/D CAL'd
indicating that calibration is
ENTER=SAVE
prompt and again at the
ENTER=SAVE
prompt. Then press [ENTER] at
prompt and again at the
.
ENTER=EXIT
prompt