Example:
Accessing Audit Trail Parameters
)00
lb±²³
Gross
60201 ó
P6020!
Audit
Trail
P6020!
OIML
00001
60203 ó
P6020#
Audit
Trail
P6020#
Cal.
00001
60204 ó
P6020$
Audit
Trail
P6020$
Setup
00001
ú
)00
lb±²³
Gross
Accessing Audit Trails
The audit trails may be accessed from the weigh mode as shown in
example – Accessing Audit Trail Parameters. An access code is not
required to view audit trail parameters.
1. Key in 60201 [SELECT] to access the OIML audit trail.
2. Key in 60203 [SELECT] to view the Calibration audit trail.
3. Key in 60204 [SELECT] to view the Setup audit trail.
4. Press [ZERO] to return to the weigh mode.
OIML Audit Trail
Changes to any of the following parameters will increment the OIML audit
trail at P60201:
•
P109 à P119
•
P122
•
P131 à P134
•
P150 à P154
•
P162, P163
•
P300 à P309
•
P410, P412
•
P420
•
P600 à P646
•
P800 à P820
•
P989 à P4999
•
P61101 à P61140 Calibration & Linearization
•
P65001, P65002
Calibration Audit Trail
Any changes to the existing calibration will increment the Calibration audit
trail at P60203. This includes any changes to P60101 à P61140 of the
information parameters.
Setup Audit Trail
Changes to any of the setup mode parameters will increment the Setup
audit trail at P60204.
D
S
ATA
TORAGE
The DSD feature of the 560 and660 Series controllers provides a means
of recording weight data in a secure database structure that ensures data
integrity. The database option is required for DSD usage. Information can
Scale Setup
Return to Zero
Units
Calibration & Custom Units
Negative Tare Enable, Tare Rounding Enable
Selectable Modes
OIML Enable, Preset Enable
Standard VF Display Mode
Rename Parameters
Key Functions
Custom Transmit
Default All, Default –Cal
D
EVICE
60 Series Technical Reference Manual
Legal-For-Trade 6-13
(DSD)