GSE 60 SERIES Technical Reference Manual page 550

Programmable process controllers
Table of Contents

Advertisement

14-10 Chapter 14
C
P
ALIBRATION
ARAMETER
P6113&
-CAL-
FACT4
P6113*
-CAL-
WGHT5
P6113(
-CAL-
FACT5
Parameters 61150 à 61160 refer to known load cell calibration values.
P6115)
#0fLC
P6115!
LC #1
FSmVv
P6115@
LC #2
FSmVv
P6115#
LC #3
FSmVv
P6115$
LC #4
FSmVv
P6115%
LC #5
FSmVv
P6115^
LC #6
FSmVv
P6115&
LC #7
FSmVv
P6115*
LC #8
FSmVv
P6115(
LC FS
100.0
P6116)
Avg.
LCmVv
Parameters 61200 à 61207 refer to analog output calibration values.
P6120)
AnOut
P6120!
V Z —
xxxxx
P6120@
V G —
xxxxx
P6120#
0mAZ—
xxxxx
P6120$
0mAG—
xxxxx
P6120%
4mAZ—
xxxxx
P6120^
4mAG—
xxxxx
P6120&
SN
12345
GSE Scale Systems
Displays the calibration factor for weights within the range of the 4
calibration point.
Displays the weight (in default units per P150) used for the 5
calibration point.
Displays the calibration factor for weights within the range of the 5
calibration point.
Displays the number of load cells entered during the last "known load cell output"
calibration. This value is reset to zero (0) if the "known load cell output" was not the
0
last calibration method used.
Displays the full scale mV/V output entered for load cell #1.
Displays the full scale mV/V output entered for load cell #2 (if specified).
Displays the full scale mV/V output entered for load cell #3 (if specified).
Displays the full scale mV/V output entered for load cell #4 (if specified).
Displays the full scale mV/V output entered for load cell #5 (if specified).
Displays the full scale mV/V output entered for load cell #6 (if specified).
Displays the full scale mV/V output entered for load cell #7 (if specified).
Displays the full scale mV/V output entered for load cell #8 (if specified).
Displays the full scale capacity entered for the individual load cells.
Displays the calculated average full scale mV/V for all values entered in P61151 à
P61158.
Used to enter the analog output number for P61201 à P61207. For example, key
in 2 [ENTER], then access P61201 à P61207 to view calibration data for analog
1
output #2.
Sets the zero reference for the 0-10VDC analog output. Enter the value that results
in exactly 0.00VDC output (range is 0 à 15000).
Sets the 10V reference for the 0-10VDC analog output. Enter the value that results
in exactly 10.00VDC output (range is 50000à 65535).
Sets the zero reference for the 0-20mA analog output. Enter the value that results
in exactly 0.00mA output (range is 0 à 15000).
Sets the 20mA reference for the 0-20mA analog output. Enter the value that results
in exactly 20.00mA output (range is 50000à 65535).
Sets the zero reference for the 4-20mA analog output. Enter the value that results
in exactly 4.00mA output (range is 0 à 15000).
Sets the 20mA reference for the 0-20mA analog output. Enter the value that results
in exactly 20.00mA output (range is 50000à 65535).
:
Displays the analog output's serial number. This number should be entered at the
time of installation for future reference.
D
ESCRIPTION
th
linearization
th
linearization
th
linearization