Post-Repair Procedures
Post-Repair Procedures
Post-Repair Procedures
Table 15-1
after an assembly replacement.
After an assembly is replaced, find the assembly that has been replaced in the
left-hand column, and then perform the recommended adjustment and/or
performance verification test. Making the recommended adjustments and/or
performance verification tests doesn't guarantee all published specifications
are being met. Only a full instrument calibration will do this.
Refer to the analyzer user's guide for information on instrument warm-up
NOTE
before performing any of the procedures listed in this chapter.
The following procedures are also included in the
page
—
—
—
—
—
—
lists the adjustments and performance verification tests needed
567.
"Calibration Data Backup and Restore"
"BIOS Settings Verification"
"End User License Agreement (EULA) Acceptance (Software versions ≤
A.06.08)"
"End User License Agreement (EULA) Acceptance (Software versions ≥
A.07.00)"
"FPGA Synchronization"
"Programming Model and Serial Numbers"
562
"Additional Tasks" on