Functional Tests

Frequency Response (Flatness)

Frequency Response (Flatness)
Test Limits
20 Hz to 3.6 GHz
> 3.6 GHz to 8.4 GHz
> 8.4 GHz to 13.6 GHz
> 13.6 GHz to 26.5 GHz
The frequency response test measures the signal analyzer's amplitude error as
a function of the tuned frequency. Measurements are made ranging from 50
MHz to the maximum frequency range of your analyzer. The signal source
amplitude is measured with a power meter to eliminate error due to source
flatness. The measured value is normalized to 50 MHz.
Item
Adapter, Type N (m) to
3.5 mm (m)
Adapter,
3.5 mm (f) to 3.5 mm (f)
Cable, 3.5 mm, 1 meter
Cables, BNC 120 cm
Power Meter
Power Sensor
Power Splitter
Synthesized Sweeper
Frequency Range
Critical Specifications
(for this test)
Frequency: 10 MHz to 18 GHz
VSWR: < 1.1:1
Frequency: 10 MHz to 26.5 GHz
VSWR: < 1.1:1
Frequency: 10 MHz to 26.5 GHz
VSWR: < 1.4:1
Frequency: 10 MHz
Compatible with power sensor
Frequency Range: 50 MHz to 26.5 GHz
Frequency Range: 50 MHz to 26.5 GHz
Tracking between ports: < 0.25 dB
Frequency Range: 50 MHz to 26 GHz
616
Limit Relative to 50 MHz
±1.5 dB
±2.5 dB
±3.0 dB
±3.5 dB
Recommended
Keysight Model
1250-1743
83059B
11500E
10503A
E4418B
8485A
11667B
PSG