Front End Control Troubleshooting
A15 Front End Control Assembly Troubleshooting
Power Supply Verification
Looking at the test board, verify power supply green LED's DS5, DS6, DS7 and
DS8 located under the J2 connector are turned on. Additionally red LED's DS9
and DS10 should also be on. The power supply voltages are listed in the table
below.
Table 5-3
If any of the power supply LED's are not turned on, see
and Initialization

Input Attenuator A Control Logic Verification

Press AMPTD, Attenuation 0 dB on the analyzer.
Attenuation LED's DS15-20 on the test board should be off.
When the input attenuation is changed from 0 dB to 2 dB, the 2 dB Step LED
DS15 should illuminate. When the input attenuation is changed from 2 dB to 4
dB, DS15 and DS16 should be illuminated as per
Table 5-4
If this is incorrect, the most probable cause is the A15, Front End Control
board.
The analyzer default setting for RF Coupling is AC mode. Verify the AC Select
LED DS14 is illuminated at this time. Press Input/Output, RF Input, RF
Coupling, AC/DC to DC on the analyzer. DS14 should turn off and DS11, DC
Select LED should illuminate. Switch the RF Coupling back to AC. If the LED's
do not illuminate correctly, the most probable cause is the A15 Front End
Control Assembly.
Power Supply Voltages
LED
Power Supply
DS5
+15 VDC
DS6
+9.0 VDC
DS7
+5.2 VDC
DS8
-15 VDC
DS9
+25 VDC (for Attenuator A)
DS10
+25 VDC (for Attenuator B)
Troubleshooting".
Input Attenuator A LED's
Attenuator Setting (dB)
0
2
4
215
Chapter 2, "Boot Up
Table
5-4.
DS15
DS16
OFF
OFF
ON
OFF
ON
ON