Post Repair Procedures - Keysight E4982A Troubleshooting Manual

Hide thumbs Also See for E4982A:
Table of Contents

Advertisement

2
Post Repair Procedure

Post Repair Procedures

Table 10 Post Repair Procedures
Replaced
Assembly or Part
A11 Source Board
A6 Receiver Board
Test Head Module
A51 DSP Board
Hard Disk Drive
52
Table 10
lists the required procedures that must be
performed after the replacement of an assembly. These are
the recommended minimum procedures to ensure that the
replacement is successfully completed.
Required Adjustment Correction
Constants (CC)
Perform the following required
adjustments using "A11 Source
Board" in Spot Adjustment of the
program.
Perform the following required
adjustments using "A6 Receiver
Board" in Spot Adjustment of the
program.
Perform the following required
adjustments using "Test Head" in
Spot Adjustment of the program.
Perform the following required
adjustments using "A51 DSP Board"
in Spot Adjustment of the program.
Perform the following required
adjustments using "HDD" in Spot
Adjustment in the program.
Verification
Troubleshooting Using Diagnostic
Test 32
Perform the following Performance
Test:
Frequency Accuracy Test
Power Level Accuracy
Impedance Measurement Accuraacy
Troubleshooting Using Diagnostic
Test 32
Perform the following Performance
Test:
Power Level Accuracy
Impedance Measurement Accuracy
Troubleshooting Using Diagnostic
Test 32
Perform the following Performance
Test:
Power Level Accuracy
Troubleshooting Using Diagnostic
Test 32
Inspect the Booting Process
Inspect the Booting Process
LCR Meter

Advertisement