Input Connector Assembly; Removal; Replacement - Keysight n9020a Service Manual

Mxa signal analyzer
Hide thumbs Also See for n9020a:
Table of Contents

Advertisement

Assembly Replacement Procedures

Input Connector Assembly

Input Connector Assembly

Removal

1. Remove the instrument outer case. Refer to the Instrument Outer
2. Remove the Front Frame Assembly. Refer to the Front Frame Assembly
3. Refer to
Figure 14-75
Input Connector Assembly Removal
4. Remove the two screws (1) that attach the Input Connector Assembly
5. To remove the gusset (5) remove the three screws (4), one on the side of

Replacement

1. Refer to
Caseremoval procedure.
removal procedure.
Figure
14-75. Disconnect the semi-rigid cable W10 from the Input
Connector Assembly (2).
(2) to the chassis. The connector assembly can now be lifted from the
chassis.
the chassis and two on the bottom.
Figure
14-75. Place the Input Connector Assembly (2) into
position in the chassis. Replace the two screws (1) to attach the Input
Connector Assembly to the chassis. Torque to 9 inch-pounds.