Analog/Digital IF Troubleshooting
40 MHz BW IF Section

A2 Analog IF Troubleshooting

There are three steps to verify the A2 Analog IF Assembly.
— Measure the input power and frequency accuracy of the 322.5 MHz signal
(from the A13 RF Front End assembly)
— Measure the output power and frequency accuracy of the 22.5 MHz signal
(to the A3 Digital IF Assembly)
— Measure the 300 MHz input LO signal (from the A16 Reference Assembly)
Verifying the 322.5 MHz Input Power
1. Turn the instrument off.
2. Remove instrument cover. Refer to
Procedures"
3. Turn on the analyzer and wait for the instrument to complete the boot up
process.
4. Press System, Alignments, Auto Align, Off.
5. Press Input/Output, RF calibrator, 50 MHz.
6. Verify the 50 MHz signal is at −25 dBm on the analyzer display by pressing
FREQ, 50 MHz, SPAN, 1 MHz, Peak Search on the analyzer. The marker
readout should be 50 MHz at −25 dBm ± 3 dB. If this reference signal is
measuring incorrectly, see
service guide.
7. To continue verifying press SPAN, Zero Span. Verify the input attenuator
on the analyzer is set to 10 dB. Look near the top of the display near the
center and verify that Atten: 10 dB is visible. If needed change the input
attenuator by pressing AMPTD, Attenuation, 10 dB on the analyzer.
in this service guide.
Chapter 4, "RF Section Troubleshooting"
255
Chapter 14, "Assembly Replacement
in this