Frequency Response (Flatness), Preamp On - Keysight n9020a Service Manual

Mxa signal analyzer
Hide thumbs Also See for n9020a:
Table of Contents

Advertisement

Functional Tests

Frequency Response (Flatness), Preamp On

Frequency Response (Flatness), Preamp On
Test Limits
100 kHz to 3.6 GHz
> 3.6 GHz to 8.4 GHz
> 8.4 GHz to 13.6 GHz
> 13.6 GHz to 26.5 GHz
The frequency response test, with preamplifier on, measures the signal
analyzer's amplitude error as a function of the tuned frequency. Measurements
are made ranging from 50 MHz to the maximum frequency range of the
preamp. The signal source amplitude is measured with a power meter to
eliminate error due to source flatness. The measured value is normalized to 50
MHz.
Item
Adapter, Type N(m) to 3.5
mm(m)
Adapter,
3.5 mm(f) to 3.5 mm(f)
Cable, 3.5 mm, 1 meter
Cables, BNC 120 cm
Power Meter
Attenuator, 30 dB Fixed
Power Sensor
Power Splitter
Synthesized Sweeper
Frequency Range
Critical Specifications
(for this test)
Frequency: 10 MHz to 18 GHz
VSWR: < 1.1:1
Frequency: 10 MHz to 26.5 GHz
VSWR: < 1.1:1
Frequency: 10 MHz to 26.5 GHz
VSWR: < 1.4:1
Frequency: 10 MHz
Compatible with power sensor
Frequency: 50 MHz
Accuracy: ±0.05 dB
Frequency Range: 50 MHz to 26.5 GHz
Amplitude Range: –65 dBm to –55 dBm
Frequency Range: 50 MHz to 26.5 GHz
Tracking between ports: < 0.25 dB
Frequency Range: 50 MHz to 26.5 GHz
620
Limit Relative to 50 MHz
±2.0 dB
±3.0 dB
±3.5 dB
±4.0 dB
Recommended
Keysight Model
1250-1743
83059B
11500E
10503A
E4418B
11708A
8485D
11667B
PSG

Advertisement

Table of Contents
loading

This manual is also suitable for:

X seriesN9020a

Table of Contents