Assembly Replacement Procedures
Analog I.F. (AIF)/Digital I.F. (DIF) Assembly
Replacement
1. Slide the AIF/DIF assembly into the slot at the rear of the instrument and
push on the assembly to mate the connectors to the midplane assembly.
2. Refer to
Figure
AIF assembly.
3. Refer to
Figure 14-61
the bottom of the DIF assembly.
4. Replace the rear panel. Refer to the Rear Panelreplacement procedure.
5. Replace the instrument outer case. Refer to the Instrument Outer
Casereplacement procedure.
14-63. Replace cables W13, W14, and W15 or W37 to the
and
Figure
14-62. Replace the appropriate cables to
515