RF Section Troubleshooting
RF Section Description
RF Section Description
Purpose
This section covers only those optional frequency ranges listed below for the
N9020A, Signal Analyzer.
— Option 503, 3.6 GHz Frequency Range
— Option 508, 8.4 GHz Frequency Range
— Option 513, 13.6 GHz Frequency Range
— Option 526, 26.5 GHz Frequency Range
The RF input signal can be routed through three different front end signal
paths. These various paths are switched in and out based on where the signal
analyzer center frequency is tuned.
1. Low band RF path is used when the input frequency is ≤ 3600 MHz
2. High band RF path #1 is used when the input frequency is > 3.6 GHz to
3. High band RF path #2 is used when the input frequency is > 13.6 GHz to
The RF section is designed to convert RF input signals to an intermediate
frequency (IF) that is present at the RF output of the A13 Front End assembly.
This IF depends on installed options and the IF path selected. This chapter will
assume the default operation in Spectrum Analyzer mode. Therefore, the IF will
be 322.5 MHz.
Table 4-1
IF Path
10 MHz (standard)
25 MHz (Option B25)
40 MHz (Option B40)
13.6 GHz
26.5 GHz
IF
322.5 MHz
322.5 MHz
250 MHz
Mode
Spectrum Analyzer or
IQ Analyzer
IQ Analyzer
IQ Analyzer
138