Troubleshooting A Low Band Problem - Keysight n9020a Service Manual

Mxa signal analyzer
Hide thumbs Also See for n9020a:
Table of Contents

Advertisement

RF Section Troubleshooting
Troubleshooting

Troubleshooting a Low Band Problem

1. Reference Assembly Verification
Remove cable W19 from A9 Input Attenuator A (1) and measure the 50 MHz
calibrator signal on the cable end with a functioning Spectrum Analyzer. Refer
to
Figure
4-5.
Figure 4-5
W19 Location
151