Assembly Replacement Procedures
Instrument Outer Case

Option PRC

Removal
1. Disconnect the instrument from ac power.
Refer to
on each side) (1) that hold the bail handle (2) to the front frame.
Figure 14-4
Bail Handle Removal
2. Using the T-20 driver, remove the four screws (two on each side) (6) that
hold the strap handle plugs (5) in place.
Figure
14-4. Using the T-20 driver, remove the four screws (two
442