Functional Tests
Frequency Response (Flatness)
Figure 16-5
Frequency Response Test Setup

Procedure

1. Zero and calibrate the power meter and power sensor as described in the
2. Configure the equipment as shown in
Connect the power splitter to the signal analyzer input using the appropriate
NOTE
adapter. Do not use a cable.
3. Assure the signal analyzer's built-in auto alignment has been performed
4. Press Mode, Spectrum Analyzer, Mode Preset on the signal analyzer, and
5. Set up the synthesized sweeper by pressing:
6. Set up the signal analyzer by pressing:
power meter operation manual.
within the last 24 hours.
press Preset on the synthesized sweeper.
CW, 50, MHz
Power level, –4, dBm
Input/Output, More, Freq Ref In, External
FREQ Channel, Center Freq, 50, MHz
SPAN X Scale, Span, 50, kHz
AMPTD Y Scale, Ref Level, 0, dBm
Figure
16-5.