Assembly Replacement Procedures
Instrument Outer Case
Instrument Outer Case
If the instrument is placed on its face during any of the following procedures,
CAUTION
be sure to use the front panel protective cover (MP50), a soft surface, or soft
cloth to avoid damage to the front panel, keys, or input connector.
If the analyzer has Option PRC, Portable Configuration, refer to the
NOTE
PRC"
Standard Instrument
Removal
1. Disconnect the instrument from ac power.
2. Refer to
3. Using the T-20 driver, remove the four screws (including washers) (3)
4. Pull the instrument cover (5) off towards the rear of the instrument.
section on
page 442
Figure
14-3. Using the T-20 driver, remove the four screws (two
on each side) (1) that attach the handle strap (2) on each side of the
instrument.
(two on each side) that hold the rear feet (4) (two on each side) in place.
to remove the outer case.
440