Summary of Contents for Keysight X Series

  • Page 1 Keysight X-Series Signal Analyzers Option SF2 Security Features, Prohibit Saving Results Notice: This document contains references to Agilent. Please note that Agilent’s Test and Measurement business has become Keysight Technologies. For more information, go to www.keysight.com. Installation Note...
  • Page 2 The information contained in this document is subject to change without notice. Keysight Technologies makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose.
  • Page 3 MXA N9020A PXA N9030A UXA N9040B UXA N9041B Serial Numbers: To Be Performed By: (X) Keysight Service Center (X) Personnel Qualified by Keysight (X ) Customer Estimated Installation Time: 1.0 Hours Estimated Adjustment and Verification Time: 0 Hours Introduction This installation note explains how to install Option SF2 on an X-Series analyzer. Option SF2 provides changes to the Softkeys and SCPI eliminating the ability to Save or Recall user data.
  • Page 4 “Instrument S/W Revision”. The latest revision of the X-Series Signal Analyzers software may be downloaded from http://www.keysight.com/find/xseries_software If a software update is required, perform the software update BEFORE following the procedure in this document for redeeming and installing the option upgrade license key.
  • Page 5: Installation Procedure

    Option SF2, Security Features, Prohibit Saving Results Installation Procedure 1. Update the instrument software to A.14.00 or later. Upon reboot assure instrument is functioning correctly and passes all auto alignments. 2. For Windows 7: Log out of the default user (instrument) by selecting Start, Log Off. For Windows 10: Log out of the default user (instrument): - Select the Start icon - Select the Change account settings icon...
  • Page 6 Option SF2, Security Features, Prohibit Saving Results 5. On the Windows desktop, click the Configure Applications shortcut. When the window appears click the Deselect All button. Click OK. This step required to assure there will be minimal applications preloaded at power up. 6.
  • Page 7: License Installation Procedure Over Usb

    USB Storage Device Configuration Menu 7. The signal analyzer will automatically consume the License File. (This may take a few minutes) When the License File is consumed the Keysight License Manager will display a “Successful License Installation” message as shown in...
  • Page 8 Option SF2, Security Features, Prohibit Saving Results Alternate Installation Procedure The License File can be manually installed over USB or LAN by placing the license file in the following folder on the signal analyzer C:\Program Files\Agilent\licensing Verify the Installation and Instrument Software Version 1.
  • Page 9 None For assistance, contact your nearest Keysight Technologies Sales and Service Office. To find your local Keysight office access the following URL, or if in the United States, call the following telephone number: http://www.keysight.com/find/assist 1-800-829-4444 (8 am - 8 pm ET, Monday - Friday)
  • Page 10 This information is subject to change without notice. © Keysight Technologies 2012-2018 Edition 1, April 2018 N9020-90234 www.keysight.com...