  • Page 1 Keysight X-Series Signal Analyzers Additional Removable Disk Drive Installation Note...
  • Page 2 The information contained in this document is subject to change without notice. Keysight Technologies makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose.
  • Page 3: Installation Note N9020-90204

    Additional Removable Disk Drive Additional Removable Disk Drive This document provides detailed instructions for the use and maintenance of an additional disk drive in an X-Series signal analyzer. Please be sure to read this entire document before attempting to put this additional disk drive into use. Contents Quantity Description...
  • Page 4: Table Of Contents

    Additional Removable Disk Drive What you will find in this User’s Guide Overview page 5 Disk Drive Interconnect Life Expectancy page 5 Putting the Additional Disk Drive Into Service page 6 Initializing the Instrument Operating System page 7 Calibration Data Backup and Restore page 9 Backup Calibration Data page 9...
  • Page 5: Overview

    A7 Midplane board assembly. Should an incorrect version of the X Series calibration file(s) be installed in an instrument, no error messages will be displayed. The instrument may however perform outside of the warranted specifications for that instrument.
  • Page 6: Putting The Additional Disk Drive Into Service

    Additional Removable Disk Drive Putting the Additional Disk Drive Into Service When ordered at the time of an instrument purchase, the additional disk drive will have the following already installed on it: 1. The currently shipping version of instrument software that is also installed in the instrument. So no instrument software updating will be required at this time.
  • Page 7: Initializing The Instrument Operating System

    The instrument may require more than 5 minutes to power-on. This is affected by the operating system initial start-up requirements. 2. The Keysight Technologies screen appears followed by a screen that automatically loads the operating system. 3. The Welcome dialog box as shown in...
  • Page 8 Additional Removable Disk Drive Figure 3 Anti-Virus Message If you do not check the “Do not show this message again” check box, this message will be displayed each time the instrument is turned on. No application will start while this message is displayed. Before continuing, make sure that you carefully read the Anti-Virus message and determine what action is appropriate.
  • Page 9: Calibration Data Backup And Restore

    Additional Removable Disk Drive Calibration Data Backup and Restore In order to maintain the same level of instrument performance and measurement accuracy with all disk drives used in any particular instrument, the instrument specific calibration data will occasionally need to be transferred from one disk drive to another. The following procedures will outline the steps required to do this.
  • Page 10: Restore Calibration Data

    Additional Removable Disk Drive Restore Calibration Data 1. With the instrument turned on and the signal analyzer application running as normal, connect a USB mouse to one of the available instrument USB ports. 2. Connect the USB storage device that the calibration data was previously saved on to one of the available instrument USB ports.
  • Page 11: Disk Drive Removal And Installation

    Additional Removable Disk Drive Disk Drive Removal and Installation Electrostatic discharge (ESD) can damage or destroy electronic components. All work on electronic assemblies should be performed at a static-safe workstation. Refer to the documentation that pertains to your instrument for information about static-safe workstations and ordering static-safe accessories.
  • Page 12: Software Updates

    The software revision can be found on this screen as the: Instrument S/W Revision Updating the Instrument Software: If an instrument software update is required, the latest revision of the X-Series signal analyzer software may be downloaded from: http://www.keysight.com/find/xseries_software Installation Note N9020-90204...
  • Page 13: Instrument Security Information

    Instrument Security Information Information on the X-Series Signal Analyzer instrument security features and the instrument volatility can be found at: http://www.keysight.com/find/security Secure Usage Scenarios The following information is meant to serve as a guideline when using the removable disk drive feature in an X-Series Signal Analyzer along with an additional disk drive option.
  • Page 14: Removing An Inoperative Instrument From A Secure Environment

    5. If the instrument failure persists with the additional disk drive the instrument should be sent in for repair to an Keysight Authorized service facility, and the rest of this procedure can be skipped.
  • Page 15 This information is subject to change without notice. © Keysight Technologies 2012-2015 Edition 1, May 2015 www.keysight.com...