Bbiq Input Frequency Response (Option Bba Only) - Keysight n9020a Service Manual

Mxa signal analyzer
Hide thumbs Also See for n9020a:
Table of Contents

Advertisement

Functional Tests

BBIQ Input Frequency Response (Option BBA only)

BBIQ Input Frequency Response
(Option BBA only)
Test Limits
Frequency Range
250 kHz to 10 MHz (Standard)
250 kHz to 25 MHz (Option B25)
250 kHz to 40 MHz (Option S40 or SU1)
This procedure tests the amplitude response of each of the BBIQ inputs versus
frequency. For each input, a reference measurement is made at 250 kHz and
then subsequent measurements are made at several frequencies up to the
maximum IQ bandwidth. The amplitude difference between the measurement
at 250 kHz and subsequent frequencies is calculated and compared to the test
limit.
RF Signal Generator
Cable, BNC
Cable, BNC, 120 cm
50 Ohm Termination
Frequency Range
Item
Frequency: 250 kHz to 40 MHz
Amplitude: ~–10 dBm
Flatness: < ±0.75 dB
Connectors: BNC male
Length: 23 cm (9 in) max
Connectors: BNC male
Frequency: 250 kHz to 40 MHz
Connector: BNC male
Frequency: 250 kHz to 40 MHz
628
Limit Relative to 250 kHz
Limit Relative to 250 kHz
± 1.75 dB
± 1.75 dB
± 1.75 dB
Critical Specifications
Recommended
Keysight Model
E4438C
10502A
10503A
11593A

Advertisement

Table of Contents
loading

This manual is also suitable for:

X seriesN9020a