Procedure - Keysight n9020a Service Manual

Mxa signal analyzer
Hide thumbs Also See for n9020a:
Table of Contents

Advertisement

Functional Tests
Amplitude Accuracy at 50 MHz
Figure 16-4
Amplitude Accuracy Test Setup

Procedure

1. Zero and calibrate the power meter.
2. Configure equipment as shown in
To minimize stress on the test equipment connections, support the power
CAUTION
sensor.
3. If the auto alignment for the analyzer has not been performed within the
4. Press Mode, Spectrum Analyzer, Mode Preset on the analyzer.
5. Set up the synthesized sweeper by pressing:
6. Set up the signal analyzer by pressing:
connected directly to the signal analyzer input through the adapter.
past 24 hours, press System, Alignments, Align Now, All to perform the
auto alignment routine.
CW, 50, MHz
Power Level, -4, dBm
RF (On)
Input/Output, More, Freq Ref In, External
FREQ Channel, Center Freq, 50, MHz
SPAN X Scale, 2, MHz
AMPTD Y Scale, Attenuation, Mech Atten, 10, dB
Figure
16-4, with the power splitter
613