Functional Tests
Frequency Readout Accuracy
Frequency Readout Accuracy
Test Limits
Frequency Readout Accuracy is equivalent to the following equation:
±
(
See results table for actual values.
The frequency readout accuracy is measured in several spans and center
frequencies that allow both internal analyzer synthesizer modes and prefilter
bandwidths to be tested. Frequency reference error is eliminated by using the
same frequency standard for the analyzer and signal source.
Item
Adapter, Type-N (m), to
3.5 mm (f)
Adapter, 3.5 mm (f) to
3.5 mm (f)
Cable, 3.5 mm,
1 meter
Cable, BNC, 120 cm
Synthesized Sweeper
Figure 16-2
Frequency Readout Accuracy Test Setup
Procedure
1. Configure the equipment as shown in
×
×
0.25% span
+
5%
Critical Specification
Frequency: 10 MHz to 1.51 GHz
VSWR: < 1.1:1
Frequency: 10 MHz to 1.51 GHz
VSWR: < 1.1:1
Frequency: 10 MHz to 1.51 GHz
VSWR: < 1.4:1
Frequency: 10 MHz
Frequency: 10 MHz to 1.51 GHz
built-in auto alignment has been performed within the past 24 hours.
×
RBW
+
2 Hz
+
0.5 horizontal resolution
(for this test)
Figure
607
)
Recommended
Keysight Model
1250-1744
83059B
11500E
10503A
PSG
16-2. Confirm the analyzer's