Viewing An Intensity Scan - Thermo Scientific Accela PDA Hardware Manual

Table of Contents

Advertisement

4
Xcalibur Diagnostics for the PDA Detector
Monitoring Lamp Performance from the Xcalibur Data System

Viewing an Intensity Scan

40
Accela PDA Detector Hardware Manual
3. Save the display method as follows:
a. In the menu bar, choose Accela PDA > Direct Control.
Figure 31
shows the warning message that appears.
Figure 31. Accela PDA Diode/Intensity Warning dialog box
b. In the Accela PDA Diode/Intensity Warning dialog box, click Yes.
The Save Accela PDA Display Method dialog box appears.
c. In the File name box, type a file name.
d. Click Save.
The data system saves the file with an .spda file extension.
Note The .spda file extension is a special file extension used for all method files
based on Diode/Intensity units. These methods cannot be used in the Sequence
Setup portion of Xcalibur. You can only load files with the .spda extension in the
Direct Control dialog box. They are not tracked by the data system's auditing
feature.
To view an intensity scan
1. From the menu bar of the Accela PDA view in the Instrument Setup window, choose
Accela PDA > Direct Control.
The Accela PDA Direct Control dialog box appears.
2. Click the Display tab.
The Display page appears (see
3. In the Control area, click Load Method.
The Open Display Method dialog box appears.
Figure 32
on
page
41).