6. Depending on the current configuration of the instrument, do one of the following:
• If you are adding the Accela PDA Detector to a current configuration for your Accela
LC stack, leave the selection in the Instrument type list as Accela.
• If you want to erase the configuration for a previously configured Accela LC stack,
select Accela in the Instrument type list and click OK in the message dialog box that
appears.
• If you want to change the instrument type to Accela, select Accela in the Instrument
type list and click OK in the message dialog box that appears.
7. Click Configure.
The Accela dialog box appears.
8. In the Available modules pane, double-click the Detector button.
The Analog icon appears in the Configured modules pane.
9. To add the PDA detector to the instrument configuration:
a. Double-click the Analog button.
The Detector Configuration dialog box appears (see
Figure 25. Detector Configuration dialog box
b. In the Detector Name box, type a name to identify the Accela PDA Detector.
When you open the Instrument window, this name appears in the Analysis Channel
list.
c. In the Detector Model list, select PDA.
d. In the Acquisition Source list, select PDA Plus ACQ.
e. Leave the Y-Axis units set to the default of mAU and the Y-Axis Multiplier set to the
default of 0.001.
f.
Click
.
The Accela PDA Configuration dialog box appears (see
ChromQuest Data System Instrument Configuration
Figure
25).
Figure
26).
Accela PDA Detector Hardware Manual
3
Configuration
31