Figure 64. Wavelength File area with the Holmium12 Wavelength Calibration File
selected
4. In the Wavelength area, click Execute.
A message box appears (see
Figure 65. Calibration preconditions
5. If all of the preconditions are met, click OK.
The data system collects a background spectrum, which it uses to remove the absorbance
contribution of the mobile phase. When the background collection is complete, another
message box appears (see
5
ChromQuest Diagnostics for the PDA Detector
Calibrating the PDA Detector from the ChromQuest Data System
Figure
65).
Figure
66).
Accela PDA Detector Hardware Manual
77