Thermo Scientific Accela PDA Hardware Manual page 84

Table of Contents

Advertisement

5
ChromQuest Diagnostics for the PDA Detector
Controlling the Lamps from the ChromQuest Data System
64
Accela PDA Detector Hardware Manual
3. Record the status and usage of each lamp.
Note Avoid indiscriminately clicking either Reset button. Click them only after you
replace their associated lamp with a new one.
Use the Lamps page in the ChromQuest data system for lamp maintenance and control.
Three direct control buttons for each lamp are available on this page: On, Off, and Reset.
To use the lamp controls on the Lamps page
• In the D2 Lamp area, click On to ignite the deuterium lamp.
The Status readback displays Starting during the 10-second ignition period, and then it
changes to On. If there is a problem with the lamp, the Status readback displays Failed.
• In the W Lamp area, click On to turn on the tungsten lamp.
The tungsten lamp turns on immediately. If there is a problem with the lamp, the Status
readback displays Failed.
• Click Reset to reset the stored total run time for the associated lamp to zero and update
the Last Reset readback to the current date and time. After you replace a lamp, reset its
lamp usage hours to zero.