Xcalibur Diagnostics for the PDA Detector
Calibrating the PDA Detector from the Xcalibur Data System
48
Accela PDA Detector Hardware Manual
10. Click Finish.
The detector stores the calibration results. The date and time of the calibration appears
in the Dark Current Calibration area of the Calibration page following Last Calibration
(see
Figure
39).
Figure 39. Calibration page with dark current calibration results