Retrieving, Viewing, Printing, And Clearing The Event Log - Thermo Scientific Accela PDA Hardware Manual

Table of Contents

Advertisement

Figure 51. Final page of the Wavelength Calibration wizard

Retrieving, Viewing, Printing, and Clearing the Event Log

Thermo Scientific
13. On page 8 of the Wavelength Calibration wizard (see
The final page of the Wavelength Calibration wizard appears (see
14. Click Finish to complete the calibration.
The detector saves the calibration information. The date and time of the calibration
appear in the Wavelength Calibration area of the Calibration page (see
page
49).
Detector errors and major detector events, such as power-on self-tests (POSTs), are logged to
a dedicated area in the memory of the detector. These messages are created as part of the
normal operation of the detector and can be helpful when you are attempting to troubleshoot
communications problems.
The log can hold up to 100 events. When the log is full, the newest entry replaces the oldest
entry. To keep a continuous record for your maintenance files, print out the log and clear it
periodically. The memory is protected by battery backup when the detector is turned off.
View, print, and clear the log weekly as part of your regular maintenance routine.
For a list of error messages that can appear in the event log along with suggested solutions, see
"Log Entries"
on
page
100.
To open, print, save, or clear the event log in the Xcalibur data system
1. Open the Accela PDA Direct Control dialog box (see
Control Dialog Box"
2. Click the Information tab.
The Information page appears (see
Retrieving, Viewing, Printing, and Clearing the Event Log
on
page
34).
Figure
52).
4
Xcalibur Diagnostics for the PDA Detector
Figure 48
on
page
55), click Apply.
Figure
51).
Figure 40
"Opening the Accela PDA Direct
Accela PDA Detector Hardware Manual
on
57