Resetting the Lamp Lifetime

Setting the Startup Time for the Lamps

36
Accela PDA Detector Hardware Manual
3. Click Turn On for the associated lamp.
When you turn on the deuterium lamp, its Status readback reads Starting during the
30-second ignition period, and then it changes to On. If there is a problem with either
lamp, its Status readback reads Failed.
Note The intensity of the deuterium lamp falls off very slightly over a period of time
after the lamp is turned on. Plan to wait at least one hour for the lamp to stabilize
after a cold start before collecting data in the spectral range of the deuterium lamp.
To reset the displayed lamp lifetime
1. Open the Accela PDA Direct Control dialog box (see
Control Dialog Box"
on
2. Click the Configuration tab.
The Configuration page appears (see
3. Click Reset Lifetime for the associated lamp.
The stored total run time for the associated lamp resets to zero, and the Last Lifetime
Reset readback is updated to the current date and time.
Note Avoid indiscriminately clicking the Reset buttons. Click Reset only after you
replace the associated lamp with a new one.
To set the startup time for the lamps
1. Open the Accela PDA Direct Control dialog box (see
Control Dialog Box"
on
2. Click the Configuration tab.
The Configuration page appears (see
3. In the Programmed Lamp Startup area of the Configuration page, click Change.
page
34).
Figure 28
on
page
34).
Figure 28
on
"Opening the Accela PDA Direct
page
35).
"Opening the Accela PDA Direct
page
35).
Thermo Scientific

Advertisement

Table of Contents
loading

This manual is also suitable for:

Accela pda detector

Table of Contents