Figure 45. Page 5 of the Wavelength Calibration wizard
Thermo Scientific
6. Rotate the Holmium Oxide filter wheel to position 2 as directed.
The Next button becomes active.
Note The Next button is unavailable until the filter wheel is in position 2.
7. Click Next.
Page 5 of the Wavelength Calibration Wizard appears (see
8. Do the following:
a. Observe the status screen (see
The diagnostics program waits for a few seconds for the rise time filter to equilibrate
and takes a holmium oxide scan.
b. When you see the message: Click the Next button to proceed with calibration, click
Next.
Page 6 of the Wavelength Calibration wizard appears (see
4
Xcalibur Diagnostics for the PDA Detector
Calibrating the PDA Detector from the Xcalibur Data System
Figure
Figure
45).
Accela PDA Detector Hardware Manual
45).
Figure
46).
53