Thermo Scientific Accela PDA Hardware Manual page 59

Table of Contents

Advertisement

Figure 30. Accela PDA Method page with the selection of the Diodes/Intensity option
Thermo Scientific
2. Specify the appropriate parameters as follows:
a. To set the X-axis of the spectrum on the Display page to diode numbers, in the Units
area, select the Diode/Intensity option (see
b. To make the scan parameters available, select the Collect Spectral Data check box.
The default settings for the Diode/Intensity option appear. Use these settings to view
the light intensity counts for the entire diode array.
Start Diode = 2
End Diode = 511
Diode Step = 1
c. To make all three Channels boxes available, in the Channels area, select the Three
Channels option.
d. In the Channels boxes, type the numbers for the diodes that you want to monitor.
Note The Run Length, Filter Rise Time, and Sample Rate parameters are not
applicable for a display method.
4
Xcalibur Diagnostics for the PDA Detector
Monitoring Lamp Performance from the Xcalibur Data System
Figure
30).
Accela PDA Detector Hardware Manual
Diode/Intensity
option
39