Figure 32. Display page
Thermo Scientific
4. Browse to the display method. Then click Open to load the method.
The file name of the display method (.spda) appears in the box above the Spectrum plot.
5. To start the data stream, click Start Data (see
The Spectrum and Channels plots begin to update and the Start Data button changes to
the Snapshot button.
6. Press ALT+PRINT SCREEN to save a picture of the scan to the Clipboard. Paste this
picture into a text editor such as Microsoft Word. Keep this scan for future comparisons
to see if there is degradation in light intensity. Date the printout and add it to your
maintenance records.
Start Data –or–
Snapshot button
4
Xcalibur Diagnostics for the PDA Detector
Monitoring Lamp Performance from the Xcalibur Data System
Figure
32).
Accela PDA Detector Hardware Manual
41