13-4 Data Memory for Comparator Settings

The data memory function can store 20 sets of upper and lower limit values for the comparator mode.
The reference value or tolerance value for the comparator mode can not be stored in memory.
The upper and lower limit values in memory can be recalled easily using the MODE key and used for
weighing.
The upper and lower limit values in memory can be recalled and changed.
Note
The recalled upper and lower limit values can be changed, using the digital input mode or the
weighing input mode. The digital input mode enters the upper and lower limit values using the
keys. The weighing input mode uses a sample to store the upper and lower limit values.
Storing the upper and lower limit values
To store new upper and lower limit values: Recall the stored upper and lower limit values to be changed
("C01" to "C20") . Then, change the recalled upper and lower limit values using the digital input mode or
the weighing input mode, and store the new value.
Note
While the data memory function is in use, unit selection using the MODE key is not available.
1. Press the MODE key to select a unit to be used for
storage.
2. Set the "Data memory (data)" parameter to "3".
3. Press and hold the PRINT key (for 2 seconds) until
the balance enters the upper and lower limit values
confirmation mode. The upper limit value last selected
is displayed.
4. Select the comparator number to be used, using the following keys.
RE-ZERO key ......... To increase the comparator number by one.
MODE key .............. To decrease the comparator number by one.
Each time the RE-ZERO key or MODE key is pressed, the displayed value changes as
follows:  C03
In case of 5 level comparison, the displayed value changes in order of ...⇔C03 HI blink
ing ⇔ C03 HI ⇔ C03 LO ⇔ C03 LO blinking ⇔ C04 HI blinking.
5. To change the selected upper and lower limit values:
To use the digital input mode, press the SAMPLE key to enter the digital input mode. Go to "Digital
input mode" on the next page.
To use the weighing input mode, press the SAMPLE key, then press and hold the MODE key to
enter the weighing input mode. Go to "Weighing input mode" on the next page.
HI  C03 LO  C04
HI  C04 LO
88
No.3. Comparator number