A&D GX-M Series Instruction Manual page 121

High-capacity precision balances
Hide thumbs Also See for GX-M Series:
Table of Contents

Advertisement

(3) Entering the density of a liquid
Two ways to set the density of a liquid are available in the function table, "Liquid density input (ldin)" by
entering the water temperature or by entering the density directly.
Entering the water temperature
The water temperature currently set (unit:°C, factory setting : 25°C) is
displayed.
Use the following keys to change the value. Setting range is 0.0℃ to 99.9℃, in
increments of 0.1℃. Refer to the following matrix the " The relation between
the water temperature and density".
RE-ZERO (+) key ..... The key to increase the temperature by one degree.
MODE (-) key ........ The key to decrease the temperature by one degree.
SAMPLE key .............. Move the blinking digit.
PRINT key ................... The key to store new water temperature and return to the density
CAL key ....................... The key to cancel the change and return to the density mode.
The relation between the water temperature and density
+0
+1
0.99984 0.99990 0.99994 0.99996 0.99997 0.99996 0.99994 0.99990 0.99985 0.99978
0
0.99970 0.99961 0.99949 0.99938 0.99924 0.99910 0.99894 0.99877 0.99860 0.99841
10
0.99820 0.99799 0.99777 0.99754 0.99730 0.99704 0.99678 0.99651 0.99623 0.99594
20
0.99565 0.99534 0.99503 0.99470 0.99437 0.99403 0.99368 0.99333 0.99297 0.99259
30
0.99222 0.99183 0.99144 0.99104 0.99063 0.99021 0.98979 0.98936 0.98893 0.98849
40
Entering the density directly( Ld in 1 ) .
The density currently set (unit : g / cm
displayed.
Use the following keys to change the value.
The range to set the density is 0.0000g / cm
If it is input beyond the settable range value,
display return to the input display.
RE-ZERO (+) key ..... The key to set the value of the blinking digit .(Next to 9 will be 0.)
MODE (-) key ........... The key to select the blinking digit to change the value.(Next to 0 will be 9.)
SAMPLE key .............. Move the blinking digit.
PRINT key ................... The key to store the change and return to the density mode. (Proceed to
CAL key ....................... The key to cancel the change and return to the density mode. (Proceed to
(0 is displayed after 9 )
(9 is displayed after 0 )
mode.(Proceed to Step 5)
(Proceed to Step 5)
+2
+3
3
, factory setting : 1.0000g / cm
Step 5.)
Step 5.)
+4
+5
+6
3
3
to 1.9999g / cm
.
is displayed and the
121
+7
+8
3
) is
+9
g /cm