11-8 Comparator Function

The comparison of comparators can select 3-steps or 5-steps (Cp fnc, Cp-t), and it is set to 3-steps at the
factory setting.
When 3-step comparator is set, the results of the comparison are indicated by
the display.
When 5-step comparator is set, HH is indicated by
By using GXM-04, it is possible to output the comparison result at the contact point.
There are three types of scope that can be selected as follows.
 No comparison
 Comparison when the weight data is stable or overloaded
 Continuous comparison
The conditions for comparing near zero are in six levels from "including near zero" to "± 100 digits".
"Upper limit value and lower limit value" and "reference value and tolerance range" are the
comparison standards.
There are "Digital input" and "Input by sample load" as input method for each value.
Refer to the function setting
By setting the function setting
on the result of the comparison.
3-step comparison result
Weighing value
Threshold value
Upper limit value
Upper limit
Lower limit value ≦
Lower limit
5-step comparison result
Weighing value
Threshold value
2nd
Upper limit value
Second upper limit
Upper limit value
Upper limit
Lower limit value ≦
Lower limit
2nd
Lower limit value
Second lower limit
Note
□ The comparator function in the flow measurement mode (ApF3) is compared at the factory
setting with the flow rate value. By setting CP-Frd of the Function table CPFnc to "1", it is also
possible to compare with weight value (g unit).
Cp fnc
.
Cp beep ,
it is also possible to sound an internal buzzer depending
Judgment formura
< Weighing value
Weighing value
Weighing value
Judgment formura
Weighing value
Weighing value
Weighing value
Weighing value
Weighing value
HI blinking and LL by
3-step comparison - display
Judgment
result
HI
OK
Upper limit value
LO
Lower limit value
5-step comparison - display
Judgment
result
HH
2nd
HI
Upper limit value
Upper limit value
OK
LO
Lower limit value
2nd
LL
Lower limit value
60
HI
OK
LO
blinking.
Blinking
Lit display
display
HI
OK
LO
Blinking
Lit display
display
HI
HI
OK
LO
LO
LO
on
Buzzer
control
Buzzer
control