11-3 Description Of The Class "Environment, Display"

Condition ( Cond )
This parameter is for sensitive response to the fluctuation of a mass value. Used for
powder target mass, weighing a very light sample or when quick response weighing
is required. After setting, the balance displays
This parameter is for stable weighing with slow response. Used to prevent a mass value
from drifting due to vibration or drafts. After setting, the balance displays SLOW .
Stability band width ( 5t-b )
This item controls the width to regard a mass value as a stable value. When the fluctuation per
second is less than the parameter, the balance displays the stabilization indicator and outputs or
stores the data by function setting(dout, dataA etc.) The parameter influences the "Auto print mode".
Also, the minimum display being displayed is 1 digit.
Ex. If 0.1 g display is selected by pressing the SAMPLE key on the GX-10002M, 0.1 g is 1 digit.
This parameter is used for sensitive response of the stabilization indicator. Used for
exact weighing.
This parameter ignores slight fluctuations of a mass value. Used to prevent a mass
value from drifting due to vibration or drafts
Hold function ( Hold ) (Animal weighing mode)
This function is used to weigh a moving object such as an animal. When the weighing data is over the
weighing range from zero and the display fluctuation is within the stabilization range for a fixed period of
averaging time, the processing indicator illuminates and the balance displays the average weight of the
animal. When the animal or sample is removed from the weighing pan, the display returns to zero
automatically. This function is available only when the hold function parameter is set to "1" (the animal
mode indicator
HOLD illuminates) and any weighing unit other than the counting mode is selected. The stabilization
range and averaging time are set in "Condition (Cond)" and "Stability band width (5t-b)".
Weighing range
0.01g model
2.00g or more
0.1g model
20.0g or more
* Animal container kit (GXK-12) can be installed.
Zero tracking ( trc )
This function tracks zero point drift caused by changes in the environment and stabilizes the zero
point. When the weighing data is only a few digits, turn the function off for accurate weighing.
trc 0
The tracking function is not used. Used for weighing a very light sample.
trc 1
The normal tracking function is used.(±1 digit / 1 second)
trc 2
The strong tracking function is used. (±1 digit / 0.5 second)
trc 3
The very strong tracking function is used. (±2 digit / 0.2 second)
Averaging time
Cond 0
2
sec.
(Efficiency priority)
Cond 1
4
sec.
8
Cond 2
sec.(Exact priority)
56
FAST .
Stabilation range
5t-b 0
Lesser
5t-b 1
5t-b 2
Greater
6.25%
12.5%
16.7%