Automatic Self Calibration For The Gx-M Series - A&D GX-M Series Instruction Manual

High-capacity precision balances
Hide thumbs Also See for GX-M Series:
Table of Contents

Advertisement

Model
GX-8202M,
GF-8202M
GX-8202MD,
GF-8202MD
GX-10202M,
GF-10202M
GX-12001M,
GF-12001M
GX-22001M,
GF-22001M
GX-32001M,
GF-32001M
GX-32001MD, GF-32001MD 5kg,
Display

9-1 Automatic Self Calibration For The GX-M Series

This function automatically calibrates the balance using the internal weight due to a temperature
change in the ambient environment, time set or interval time. It can function even when the balance's
display is in off. If GLP output in function table is selected, the balance outputs the sensitivity
adjustment report or store the data in memory after calibration.
In the auto calibration mode, either the temperature change (CFnc 0), the setting time (CFnc 1), or the
interval time (CFnc 3) can be set with the function setting CFnc.
For the setting time, the three function setting of
Interval time can be set from 0.5h to 24h with function setting
Caution
If something is on the weighing pan, the balance judges that it is in use and does not perform automatic
self calibration.
The criteria that the balance judges is in use are as follows.
0.01g models
More than 20g
To maintain the calibrated state, keep the weighing pan clear while not in use.
Note The balance can be used while the indicator blinks. But, it is recommended that to
maintain the accuracy, stop using the balance and confirm that there is nothing on the pan
and allow the balance to perform self calibration.
Depending on the setting of "10.Function Switch And Initialization" , "change prohibited" or
"changeable (usable) " can be selected.
Usable calibration weight
2kg to 8kg (1kg interval)
2kg to 8kg (1kg interval)
2kg to 10kg (1kg interval)
5kg,
10kg
5kg,
10kg,
5kg,
10kg,
10kg,
This indicator means "In process of measuring calibration data".
Do not allow vibration or drafts to affect the balance while the indicator is
displayed.
0.1g models
More than 200g
Indicates that the balance detects a change in ambient temperature and
automatic self calibration will start. If the balance is not used for a few
minutes with this indicator ( ) blinking, the balance performs automatic self
calibration. The blinking duration depends on the environment.
Indicates that the balance is measuring calibration data. Do not allow
vibration or drafts to affect the balance while this indicator is displayed. After
calibration, the balance returns to indicate the previous display.
20kg
20kg, 30kg
20kg, 30kg
,
.
37
Factory
Adjustable
setting
5kg
-99.99g to
5kg
10kg
10kg
20kg
-999.9g to
20kg
20kg
and
can be set.
range
+99.99g
+999.9g