How To Set The External Weight Value - A&D GX-M Series Instruction Manual

High-capacity precision balances
Hide thumbs Also See for GX-M Series:
Table of Contents

Advertisement

9-5 How to set the External Weight Value

When calibrating the balance or performing a calibration test, the external weight you have on hand
can be set. (Refer to "Usable calibration weight" on Page 37.)
After Cal 0 is displayed, the external weight value can be set as shown in "9-3.Calibration Using
An External weight". Or, after CC 0 is displayed, the external weight value can be set as shown in
"9-4.Calibration Test Using An External Weight".
1. After displayed Cal 0 , or after displayed CC 0 , press
the SAMPLE key.
2. Press the
RE-ZERO
weight to be used.
3. Specify the calibration weight value as follows.
SAMPLE key ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ Switches the display condition to: "All of the
ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ
ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ
ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ
RE-ZERO key ꞏꞏꞏꞏꞏꞏꞏꞏꞏ Changes the external weight value (all of the
ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ
MODE key
ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ
ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ
ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ
PRINT key ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ Registers the changed external weight value.
CAL key ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ Suspends setting. (Returns to Cal 0
ꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏꞏ
key at all digits blinking to change the
segments blinking" (calibration weight
selection mode) or "The last four digits
blinking"(value adjustment mode).
segments blinking) or changes the
adjustment range (last four digits blinking).
In the adjustment range setting, the value
becomes -9999 digits after +9999 digits.
Registered values are stored even when the
power is turned off.
or CC 0 .)
41
Calibration
Calibration test
2000. 0 0
Select
the external weight
3000. 0 0
3000. 1 2
Ex: Updated the external weight
3000.12g

Advertisement

Table of Contents
loading

Table of Contents