【Entering from repeatability of external weight】
Continue from "Step 6" in "17-1 Comparing the minimum weighing value"
7. Press the PRINT key while
1KEY in .
8. Press the SAMPLE key to display EXT MA55 .
9. Press the PRINT key to proceed as order
5TART →
READY → Weighing mode.
And display requests a first load for repeatability as
LOAD
.
10. Place the weight on the weighing pan. Processing indicator
( ) illuminates.
11. When display is stabilizing for 2 seconds after blinking
processing indicator ( ), it shows span value.
12. After that, display shows
13. Processing indicator ( ) illuminates after removing the weight
from the pan.
14. When display is stabilizing for 2 seconds after blinking
processing indicator ( ), it shows zero.
15. And
is displayed as request for second load of
LOAD
repeatability.
As following procedure, perform repeatability measurement
10 times.
16. After displaying 10th span, display proceeds as order
REMOVE
minimum weighing value.
is displayed to display
MW
in blinking.
REMoVE
and shows
End
MW
0
1
%
of
_
23.20
113
MW
KEY in
EXT MA55
5TART
READY
0.00
1
LoAD
Processing indicator illuminates
1
LoAD
by placing the weight
Processing indicator blinks when
1
LoAD
stabilized
Shows span value at
1
309.56
stabilization for 2 seconds.
REMOVE
Processing indicator illuminates
REMOVE
by removing the weight
Processing indicator blinks
REMOVE
when stabilized
Shows span value at
0.00
stabilization for 2 seconds.
LoAD
10
Shows 10th span
309.55
value
REMOVE
End
0
1
%
_
MW