Output
1. Every time pressing the
amount), "TARE"(tare).
2. Only A&D standard format, DP format and CSV format are available as output format.
Output example
ST,N ,+00045.67 g
ST,G ,+00055.90 g
ST,T ,+00010.23 g
(ST,PT ,+00010.23 g )
□ By using the "UFC function", output connection and order also can be set.
For the "UFC function", please refer to "Communication manual" which can be downloaded from
the A&D website (https://www.aandd.jp).
16-2

Example Of Using The Gross Net Tare Function

1. After setting the gross net tare function, press the ON:OFF
when nothing is on the weighing pan. "G" will be displayed on the display.
2. Place the container to be tared on the weighing pan.
3. Press the RE-ZERO key to display
the tare value is set (updated). "NET" is displayed on the display.
4. Place the object.
5. Press the PRINT key, it will output in the order of
"NET"(object), "GROSS"(total amount), "TARE"(tare).
6 Remove anything on the weighing pan and press the
key to return to the "1".
To continue weighing without changing the tare value,
remove the object only, place the next weighing object and
press the PRINT key to continue outputting.
PRINT key, it will output in the order of "NET"(object), "GROSS"(total
(A&D standard format)
NET(object)
GROSS(total amaount)
TARE(tare)
PRESET TARE(tare)
When the unit setting of the balance is PCS or %,
the
unit
"PRESET TARE" become "g" unit.
output
of
"GROSS","TARE"
NET
,
ON:OFF