Outputting the statistical results
1.
Each time the
changes: the results as selected in "Statistical function
mode output items (5taf)", and
.
When pressing the
previous item is displayed.
Note
□ When the number of data is 1, the coefficient of
variation and relative error is displayed as - - - - - - - .
□ When the average is 0, the coefficient of
variation and relative error is displayed as - - - - - - - .
□ Statistical items are indicated on the upper
left of the display using the following
symbols.
2. When pressing the
displaying the statistical result, the statistical
result is output.
Symbol
Range (Maximum – minimum)
Relative error of maximum value
Relative error of minimum value
Deleting the latest data
When the wrong data is entered, it can be deleted and excluded from
statistical calculation. Only the latest data can be deleted.
1. In the weighing mode, press the MODE
times to display
2.
Press the PRINT key to display
3.
Press the RE-ZERO key to display
4.
Press the PRINT key to delete the latest data and
exclude it from statistical calculation.
The number of data decreases by 1 when the balance
returns to the weighing mode.
MODE key is pressed, the display
SAMPLE key, the
PRINT key while
Statistical item
Sum
Maximum
Minimum
Average
Standard deviation
Coefficient of variation
1
.
,
Output the statistical result
Delete
the statistical data
Delete
the latest data
Output example
N
SUM
MAX
MIN
R
AVE
SD
CV    2.8 %
MAX%
MIN%
 
key several
.
.
100
Display
the statistical result
Function table
parameter
10
0
1000.00g
1
105.00g
2
95.00g
3
10.00g
100.00g
2.800g
5.0 %
5.0 %
Press several times