Description Of The Data Format Added To The Weighing Data - A&D MC-1000 Instruction Manual

Mass comparator balances
Hide thumbs Also See for MC-1000:
Table of Contents

Advertisement

5if type 5
CSV format
 Separates the data of A&D standard format and the unit by a comma (,).
 Outputs the unit even when the data is overloaded.
 When the ID number, data number, time and date are added, outputs the ID number, data
number, date, time and weighing data in this order and separates each item by a comma and
treats all the items as one group of data.
LAB-123, No,012, 2009/12/31, 12:34:56, ST,+1000.0000,
ID number Data number Date
S T , + 1 0 0 0 . 0 0 0 0 ,
O L , +
Note
To add the ID number, data number, time and date, the function settings must be changed.

10-7 Description of the Data Format Added to the Weighing Data

dout 5-id 1
ID number
The number to identify a specific balance.
 This format consists of seven characters excluding the terminator.
Data number dout d-no 1
This format outputs the data number just before the data is transmitted using the RS-232C interface.
 This format consists of six characters excluding the terminator.
 When CSV format ( 5if type 5 ) is selected, the period ( . ) is replaced with a comma (
Note
The data number is added only to the weighing data that is stored in memory.
dout 5-td 2
Date
 The date output order can be changed in "Clock ( Cl adj )". Outputs the year in four-digit
format.
2 0 0 9 / 1 2 / 3 1
dout 5-td 1
Time
 Outputs time in 24-hour format.
Note
When the data described above is added to the weighing data, the output is in the following
order: ID number, Data number, Date, Time and Weighing data.
Time
9 9 9 9 9 9 9 E + 1
3
or
C R L F
3
or
Weighing data
C R L F
g
9 ,
g
41
g<CR><LF>
C R L F
,
).

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mc seriesMc-6100Mc-6100sMc-1000s

Table of Contents