5if type 5
CSV format
Separates the data of A&D standard format and the unit by a comma (,).
Outputs the unit even when the data is overloaded.
When the ID number, data number, time and date are added, outputs the ID number, data
number, date, time and weighing data in this order and separates each item by a comma and
treats all the items as one group of data.
LAB-123, No,012, 2009/12/31, 12:34:56, ST,+1000.0000,
ID number Data number Date
S T , + 1 0 0 0 . 0 0 0 0 ,
O L , +
Note
To add the ID number, data number, time and date, the function settings must be changed.
10-7 Description of the Data Format Added to the Weighing Data
dout 5-id 1
ID number
The number to identify a specific balance.
This format consists of seven characters excluding the terminator.
Data number dout d-no 1
This format outputs the data number just before the data is transmitted using the RS-232C interface.
This format consists of six characters excluding the terminator.
When CSV format ( 5if type 5 ) is selected, the period ( . ) is replaced with a comma (
Note
The data number is added only to the weighing data that is stored in memory.
dout 5-td 2
Date
The date output order can be changed in "Clock ( Cl adj )". Outputs the year in four-digit
format.
2 0 0 9 / 1 2 / 3 1
dout 5-td 1
Time
Outputs time in 24-hour format.
Note
When the data described above is added to the weighing data, the output is in the following
order: ID number, Data number, Date, Time and Weighing data.
Time
9 9 9 9 9 9 9 E + 1
3
or
C R L F
3
or
Weighing data
C R L F
g
9 ,
g
41
g<CR><LF>
C R L F
,
).