Data Memory; Data Memory For Weighing Data - A&D GX-M Series Instruction Manual

High-capacity precision balances
Hide thumbs Also See for GX-M Series:
Table of Contents

Advertisement

13.Data Memory

Data memory is a function to store weighing data and calibration data in memory. The data stored
in memory are available for outputting at one time to a printer or personal computer.
The following five types of data can be stored.
Unit mass (Counting mode)
Calibration report
Internal calibration
External calibration
Calibration test report
Internal test calibration
External test calibration
Comparator setting values
Upper limit value/lower limit value only
Tare value

13-1 Data Memory For Weighing Data

Features
□ It is not necessary to connect the printer or personal computer to the balance continually, because
the balance stores the weighing data in memory.
□ By storing the weighing value in the balance, weighing operation can be performed without occupying
the printer or PC for a long time.
□ The data in memory can be displayed on the balance for confirmation.
□ Data (ID number, data number, time and date) to be added to the output data can be selected in
the function setting.
□ Up to 200 sets of weighing data including time and date can be stored in memory of the balance.
* For the unit mass storage method, refer to "6-2 Counting Mode(PCS)".
Storing the weighing data and calibration history
Note
1. Set the "Data memory ( data )" parameter to "
2. Set the "Time/Date output ( 5-td )" parameter about whether or not to add time and date.
3. The storing mode depends on the "Data output mode ( prt )" parameter setting.
When set to prt3 (stream mode), data may not be stored correctly.
*
It is also possible to change the time / date setting after storing the weighing value.
Weighing value
data 2
". Refer to "11. Function Table".
78
Up to 50 sets
Up to 200 sets
Last 50 sets
Up to 20 sets
Up to 20 sets

Advertisement

Table of Contents
loading

Table of Contents