Main Display Comparison Function - A&D GX-M Series Instruction Manual

High-capacity precision balances
Hide thumbs Also See for GX-M Series:
Table of Contents

Advertisement

11-10 Main Display Comparison Function

The main display comparison function displays the comparison results in a magnified way, on the main
portion of the display in place of the weight value.
Selecting a unit
Step
1 Press the MODE key to select a unit to be used for comparison.
Note While the main display comparison function is in use, unit selection using the MODE
key is not available.
Setting the function table
2 Press and hold the SAMPLE key (for 2 seconds) until ba5fnc of the function table is
Step
displayed, then release the key.
3 Press the SAMPLE key several times to display Cp fnc .
Step
Step
4 Press the PRINT key.
Step
5 Press the SAMPLE key several times to display Cp-b 0 .
Step
6 Press the RE-ZERO key to display Cp-b 1 .
Note To disable the main display comparison function, set the "Main display comparison
(Cp-b)" parameter to "0".
Step
7 Press the PRINT key to store the setting.
Step
8 Press the CAL key to return to the weighing mode.
Setting the comparator values
Setting the comparator values as described in the previous section.
This example uses Cp 3 (Continuous comparison, excluding "near zero").
Using the main display comparison function
Step
1 Press the RE-ZERO key to set the display to zero.
Step
2 Place a sample on the pan. The balance performs a comparison using the specified comparison
values and displays the comparison results, HI, OK or LO.
Step
3 Each time the MODE key is pressed, the balance switches between the standard display and
the main display comparison. Note that "go" appears for OK.
Notes
While the main display comparison function is in use, the processing indicator
illuminates as shown in the illustration.
If the comparison is not performed, for example, because the weight value is near
zero or unstable, the balance displays the weight value even when the main display
comparison function is used.
Even while the main display comparison function is in use, the balance re-zeroing
and data output are possible.
Only the unit selected before this function can be used.
While the main display comparison function is in use, the data memory function is
not available.
To disable the main display comparison function, set the "Main display comparison
(Cp-b)" parameter to "0".
67