A&D GX-M Series Instruction Manual page 25

High-capacity precision balances
Hide thumbs Also See for GX-M Series:
Table of Contents

Advertisement

About the operation at when power is turned on
The balance will decide the reference zero point when the power is turned on (AC adapter is
connected).
Depending on the load condition at that time, it will automatically judge whether to perform zero or tare
operation. The condition for determining which is used is "power on zero range", and when power on
zero range is exceeded, the tare subtraction operation is performed.
About re-zero operation
By pressing the
RE-ZERO
Re-zero with the
RE-ZERO
performed.
The condition for determining which is used is "zero range", and when zero range is exceeded, the tare
subtraction operation is performed.
About measurement range
For the balance, the range that can be weighed is determined by model.
The total amount (net amount + tare quantity) up to the maximum display of each model is displayed,
and when the maximum display is exceeded,
exceeded.
When in excess in negative,
Model
GX-8202M,
GF-8202M
GX-8202MD,
GF-8202MD
GX-10202M,
GF-10202M
GX-12001M,
GF-12001M
GX-22001M,
GF-22001M
GX-32001M,
GF-32001M
GX-32001MD, GF-32001MD Approx.±3kg
key, the display can be changed to zero.
key will automatically determine whether zero or tare operation is
E is displayed to indicate that the weighing range is
-E is displayed.
Power on zero
range
Approx.±800g
Approx.±800g
Approx.±1kg
Approx.±1kg
Approx.±2kg
Approx.±3kg
25
Zero range
Approx.-800g to +160g
Approx.-800g to +160g
Approx.-1kg to +200g
Approx.-1kg to +200g
Approx.-2kg to +400g
Approx.-3kg to +600g
Approx.-3kg to +600g
-E display range
Approx.-800g or less
Approx.-800g or less
Approx.-1kg or less
Approx.-1kg or less
Approx.-2kg or less
Approx.-3kg or less
Approx.-3kg or less