A&D GX-M Series Instruction Manual page 98

High-capacity precision balances
Hide thumbs Also See for GX-M Series:
Table of Contents

Advertisement

Selecting the statistical items to output
5. Press the SAMPLE key to display
6. Press the RE-ZERO key to select the output items.
In the example, output the number of data, sum, maximum,
minimum, range (maximum-minimum) and average are
selected.
Parameter
0
Number of data, sum
Number of data, sum
1
Maximum, minimum, range (maximum – minimum),
average
Number of data, sum
2
Maximum, minimum, range (maximum – minimum),
average, Standard deviation, coefficient of variation
Number of data, sum
Maximum, minimum, range (maximum –
minimum),average, Standard deviation,coefficient of
3
variation
Relative error of maximum value, relative error of minimum
value
7. Press the PRINT key to store the setting.
8. Press the CAL key to return to the weighing mode.
Selecting the unit
9. Press the MODE key to select the unit to be used for the
statistical calculation mode. In the example shown at the right,
gram (g) is selected.
Note
Selecting the unit using the MODE key is not available after the data is entered. In this case,
clear the data as described on page 99 "Cleaning the statistical data" and select the unit using
the MODE key.
When the unit used for the statistical calculation mode is to be enabled upon power-on, select
the unit in "Unit ( Unit )" of the function table beforehand.
.