Gross Net Tare Function; Preparation Of Gross Net Tare Function - A&D GX-M Series Instruction Manual

High-capacity precision balances
Hide thumbs Also See for GX-M Series:
Table of Contents

Advertisement

16.Gross Net Tare Function

Zero setting and taring can be operated separately, and data output for Gross (total amount), Net (net
amount), Tare (tare quantity) becomes possible.
When the gross net tare function is selected, the key operation is changed as follows.
Key
ON:OFF key
RE-ZERO key
In order to use the Gross Net Tare Function, it is necessary to change the "setting of the Function
table".

16-1 Preparation Of Gross Net Tare Function

To use this function, enter the Function table as follow, and set "Application Function ap fnc " to "4 " in
"Application mode apf ". To return the normal weighing mode (Factory settng), set " Application mode
apf " to "0 ".
Please set as follows.
Setting procedure
1. Press and hold the
1ba5fnc
of the function table is displayed, then release
the key.
2. Press the SAMPLE key several times to display
3. Press the PRINT key to display
4. Press the RE-ZERO key several times to
display
5. Press the PRINT key to store the setting.
6. Press the CAL key to return to the calculating display.
Operation
Zero setting (Operate as the
Tare(Operate as the TARE key)
SAMPLE key
(for 2 seconds)
.
ZERO
key)
until
ap fnc .
.
108
Press and hold
(for 2 seconds)
Press several times
Press several times
end

Advertisement

Table of Contents
loading

Table of Contents