A&D GX-M Series Instruction Manual page 105

High-capacity precision balances
Hide thumbs Also See for GX-M Series:
Table of Contents

Advertisement

Setting of flow rate unit
5. Press the SAMPLE key to display
6. Press the RE-ZERO key to set the setting value.
Parameter
g / s (gram/second)
0
g / m (gram/minute)
1
g / h (gram/hour)
2
mL / s (milliliter/second)
3
mL / m (milliliter/minute)
4
mL / h (milliliter/hour)
5
7. Press the PRINT key to store.
8. Press the CAL key to return to the calculating display.
Factory setting
Manual / automatic selection of flow calculation time Ct.
There are two ways to set flow calculation time Ct, either by automatic setting in the balance
according to the flow rate value or by manually determining a fixed value.
To switch between manual and automatic, please perform the following operation.
In factory setting, flow calculation time Ct is set to manual
input setting ( Ct AUto "oFF" ) .
1. Please perform the following operation from the
display for flow unit setting as shown in "15-1. How To Use
Flow Measurement".
2. Press the SAMPLE key to display
3. Press the RE-ZERO key to change ON/OFF .
4. Press the PRINT key to store.
5. Press the CAL key to return to the calculation display.
* If set to "OFF", refer to "How to set calculation time by
manual setting" to set the flow calculation time.
If set to "ON", refer to "How to set calculation time by
automatic setting" to set the flow calculation precision.
Contents
.
.
Ct AUto
105
0
Frd Un IT
5
3
Frd Un IT
mL
/
5
f01
mL
0.00
/
5
From setting of flow rate unit
3
Frd Un IT
mL
/
5
0
0FF
Ct AUto
1
0N
Ct AUto
End
MW fnc
f01
mL
0.00
/
5