A&D GX-M Series Instruction Manual page 115

High-capacity precision balances
Hide thumbs Also See for GX-M Series:
Table of Contents

Advertisement

19. After output,
MW
20. Press and hold the
0
1
returns to
_
MW
* In this time, when MW-CP is set to 0, the balance automatically set it to 1 (except near zero) and
enables comparing function of the minimum weighing value.
21. Press the
CAL key twice to proceed to weighing mode. Alarm function of minimum weighing
value is started.
17-2-2 Setting from the weighing mode
1. Press the MODE key in weighing mode.
After confirming
press the PRINT key.
2.
KEY in is displayed.
As following procedure, continue to Step 8
of 【Inputting the setting value directly】 or
Step 8 of【Entering from repeatability of
external weight】in "17-2-1 Setting from
function setting" to set minimum weighing
value.
17-2-3 Outputting the setting value at once
The minimum weighing value set currently and results for repeatability can be output at once.
1. Press and hold the
while
KEY in or Ext MA55 is displayed.
2. Press the REZERO key to toggle between "No" or
"Go" on OuT
display.
Select"Go"and press the
Setting values are outputted at once.
0
1
%
_
is displayed.
23.20
SAMPLE key (for 2 seconds) to store minimum weighing value. Display
%
.
23.20
KEY
display,
MW
23.20
PRINT key (for 2 seconds)
PRINT key.
Setting process is displayed
at unit display above.
KEY Input setting value
EXT Input from repeatability for the external weight
ECL Input by ELC (press and hold the MODE key)
Refer to "8-2 Self-Check-Function / Automatic Setting Of
Minimum Weight Value".
MW
KEY in
Press and hold (for 2 seconds)
Out
Batch output
End
115
0.00
KEY
MW
23.20
KEY in
EXT MA55
Out
Go
No