A&D GX-M Series Instruction Manual page 135

High-capacity precision balances
Hide thumbs Also See for GX-M Series:
Table of Contents

Advertisement

Display
5D Error
MW Error
Alternate
Error code
EC,E21 Calibration weight error (Negative value)
The calibration weight is too light. Confirm the calibration mass
value. Press the CAL key to return to the weighing mode.
Overload error
A sample beyond the balance weighing capacity has been
placed on the pan. Remove the sample from the pan.
Weighing pan Error
The mass value is too light.
Confirm that the weighing pan is properly installed and calibrate
the balance.
Sample mass error
The balance can not store the sample for the counting mode or
for the percent mode because it is too light.
Use a larger sample.
Unit mass error
The sample unit mass for the counting mode is too light. Storing
and using it for counting will cause a counting error.
Add samples to reach the specified number and press
the ..PRINT key. Pressing the
adding samples will shift the balance to the counting mode.
But, for accurate counting, be sure to add samples.
Clock battery error
The clock backup battery has been depleted.
Press any key and set the time and date. The clock and calendar
function works normally as long as the AC adapter is connected
to the balance. If this error appears frequently, contact the local
A&D dealer.
Power supply voltage fault
The voltage supplied from the AC adapter is abnormal.
Please check if the problem is the AC adapter (TB248)
supplied with the balance.
ECL repeatability
With the self-check function, the standard deviation (SD) of
repeatability due to electronically controlled load (ECL)
exceeded 50 digits.
Please revise the installation environment of the balance.
□ 5D Error
This is displayed when repeatability is displayed by ECL.
□ MW Error
This is displayed when the minimum weighing value
(reference value) by ECL is displayed.
Refer to "8-2.Self-Check-Function / Automatic Setting Of
Minimum Weight Value.
Full memory
The maximum number or stored weighing values has been
Blink
reached. In order to store more weighing values, it is
necessary to delete the data. Refer to "13.Data Memory"
Description
135