Enabling data memory function
1. Press and hold the
1ba5fnc of the function table is displayed.
2. Press the SAMPLE key several times to display dout .
3. Press the PRINT key.
4. Press the SAMPLE key several times to display data .
5. Press the RE-ZERO key to display
6. Press the PRINT key to store the setting.
7. Press the CAL key to return to the weighing mode.
Display and symbol
Number of stored data
Full memory
Data number of the
displayed weighing
value
Caution
□ When weighing data is being stored in memory, the data is output simultaneously using RS-232C interface or
USB.
□ "fUl" indicates that memory is full or the memory capacity has been reached. More data cannot
be stored unless the memory data is deleted.
□ Automatic self calibration can not be used while the interval memory mode is active.
□ Statistic calculation function can not be used when the data memory function is active.
SAMPLE
key (for 2 seconds) until
Weighing display in
Display of measured weighing
value
79
.
progress
Press and hold
(for 2 seconds)
bA5Fnc
Press several times
dout
0
OFF
dAtA
2
CAL
dATA WT/
5 if
When the volume of measured
values stored reaches its
fUl
dat
maximum,
blink in turn.